VUMX
(edit with the Customer Reassurance module)
(edit with the Customer Reassurance module)
(edit with the Customer Reassurance module)
Dedicated circuit recommended and pre-installation form required. 2 year warranty on both mobile and wall mounted generators.
Generator Specs
Line Voltage 100V/240V 50/60 Hz
AMP draw 8.5A (@100V)
Technique ranges:
X-RayTube Voltage: 60/65/70 kV
X-Ray Tube Current: 4-7mA (+/-1mAsteps)
Exposure times: 0.02-2s (21steps, R'10)
Maximum Combinations of Loading Factors
70 kV/6 mA/2 s
65 kV/7 mA /2 s
60 kV or 65 kV=7mAmax./2s max.
70 kV=6 mA max./2 s max.
Cone length: 20 CM (standard)
Total filtration: 2.2 mm Al/70 kV
Inherent Filtration: 1.2mm Al equivalent /70kV
Focal spot: 0.4 mm
Duty cycle: 1:30
Remote hand switch length: 13feet